Item type |
紀要論文 / Departmental Bulletin Paper(1) |
公開日 |
2010-02-05 |
タイトル |
|
|
タイトル |
院内PHSが電子ポケット線量計に及ぼす影響 |
言語 |
|
|
言語 |
jpn |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
電子ポケット線量計 |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
電磁波耐性 |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
院内PHS |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
携帯電話機 |
キーワード |
|
|
主題Scheme |
Other |
|
主題 |
放射線被曝 |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
departmental bulletin paper |
著者 |
西村, 裕子
千田, 浩一
佐藤, 友子
遠藤, 歩美
坂本, 睦美
星, 千春
及川, 喜弘
長尾, 眞人
|
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329921 |
|
|
姓名 |
NISHIMURA, Yuko |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329922 |
|
|
姓名 |
CHIDA, Koichi |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329923 |
|
|
姓名 |
SATO, Yuko |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329924 |
|
|
姓名 |
ENDO, Ayumi |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329925 |
|
|
姓名 |
SAKAMOTO, Mutsumi |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329926 |
|
|
姓名 |
HOSHI, Chiharu |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329927 |
|
|
姓名 |
OIKAWA, Yoshihiro |
著者別名 |
|
|
|
識別子Scheme |
WEKO |
|
|
識別子 |
329928 |
|
|
姓名 |
NAGAO, Makoto |
抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
It is well known that the electromagnetic fields emitted by cellular telephones can affect medical electronic devices, such as pacemakers. Therefore, it is recommended that cellular telephones with lower electronic field strengths be used in Japanese hospitals (hospital cellphones) The electronic pocket dosimeter (EPD) is a device that is used for real-time monitoring of radiation exposure from radiation sources such as radioisotopes and X-ray tubes. Recently, with the increased use of positron emission tomography (PET) examination, EPDs have become more important for radiation monitoring in nuclear laboratories, especially PET laboratories. Although there have been some reports that the electromagnetic fields emitted by cellular telephones (conventional cellphones) can cause abnormally high EPD values, it is not clear whether the electromagnetic fields emitted by hospital cellphones affect EPDs. Therefore, we investigated the effects on EPDs of the electromagnetic fields emitted by hospital cellphones. Two types of EPD (with more or less shielding against electromagnetic fields: high-shield and low-shield EPDs, respectively) were exposed to the fields emitted by two types of cellular telephone for 20 seconds (conventional cellphone vs. hospital cellphone). With the high-shield EPD, neither the conventional cellphone nor the hospital cellphone had any effect. With the low-shield EPD, the distance that prevented electromagnetic interference (protection distance) was identical for both types of cellphone (2.5±0.58vs.2.75±1,5cm,p=0.766). As regards the relative positions of the EPDs and cellphones, there was no significant difference in the protection distance between side-by-side and overlapping (p=0.100), although the distance tended to be greater in the overlapping position (2.0±0.82vs. 3.25±0.96 cm,p=0.100). The maximum dosage error for the low-shield EPD tended to be lower for hospital cellphones than for conventional cellphones (1,789±1,547vs. 3,868±1.321μSv), but the difference was not statistically significant (p=0.111). Although the electronic field emitted by hospital cellphones is weaker than that emitted by conventional cellphones, we found that both types produced abnormally high readings with low-shield EPDs. Therefore, we recommend the use of high-shield EPDs. |
書誌情報 |
東北大学医学部保健学科紀要
en : Bulletin of School of Health Sciences Tohoku University
巻 15,
号 1,
p. 29-34,
発行日 2006-01-31
|
ISSN |
|
|
収録物識別子タイプ |
ISSN |
|
収録物識別子 |
1348-8899 |
書誌レコードID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA11944693 |
フォーマット |
|
|
内容記述タイプ |
Other |
|
内容記述 |
application/pdf |
著者版フラグ |
|
|
出版タイプ |
VoR |
|
出版タイプResource |
http://purl.org/coar/version/c_970fb48d4fbd8a85 |
その他の言語のタイトル |
|
|
その他のタイトル |
Effects of Electromagnetic Fields Emitted from Cellular Telephones for In-hospital Use on Electronic Pocket Dosimeters |
出版者 |
|
|
出版者 |
東北大学医学部保健学科 |
資源タイプ |
|
|
内容記述タイプ |
Other |
|
内容記述 |
紀要類 |
登録日 |
|
|
日付 |
2010-02-05 |
|
日付タイプ |
Created |
公開日(投稿完了日) |
|
|
日付 |
2010-02-05 |
|
日付タイプ |
Created |
発行日 |
|
|
日付 |
2006 |
|
日付タイプ |
Created |
フォーマット |
|
|
内容記述タイプ |
Other |
|
内容記述 |
893977 bytes |
著者所属 |
|
|
値 |
名古屋大学医学部保健学科放射線技術科学専攻 / 東北大学医学部保健学科放射線技術科学専攻 / 東芝メディカル株式会社 / 東北公済病院放射線室 / 仙台赤十字病院放射線室 / 宮城県対がん協会放射線科 / 国立病院機構仙台医療センター放射線科 / 国立病院機構仙台医療センター放射線科@@@Department of Radiological Technology, Nagoya University School of Health Sciences@@@Department of Radiological Technology, School of Health Sciences, Faculty of Medicine, Tohoku University@@@Toshiba Medical Systems Corporation@@@Department of Radiology, Tohoku Kosai Hospital@@@Department of Radiology, Japanese Red Cross Sendai Hospital@@@Department of Radiology, Miyagi Cancer Society@@@Department of Radiology, Sendai Medical Center, National Hospital Organization@@@Department of Radiology, Sendai Medical Center, National Hospital Organization |
更新日 |
|
|
日付 |
2010-02-05 |
|
日付タイプ |
Created |